實務上要進行光學元件的透光率、OD量測時,光譜儀的選擇注意"規格"上的限制。
首先可以先參考在台灣一般光譜量測設備經常使用到的 Ocean Optics 的 USB2000+ 這一顆模組化光譜儀的規格:
Engineering Specifications | USB2000+ |
PHYSICAL | |
Dimensions: | 89.1 mm x 63.3 mm x 34.4 mm |
Weight: | 190 g |
DETECTOR | |
Detector: | Sony ILX511B (2048-element linear silicon CCD array) |
Detector range: | 200-1100 nm |
Pixels: | 2048 pixels |
Pixel size: | 14 µm x 200 µm |
Pixel well depth: | ~62,500 electrons |
SPECTROSCOPIC | |
Optical resolution: | ~0.1-10.0 nm FWHM (configuration dependent) |
Signal-to-noise ratio: | 250:1 (full signal) |
A/D resolution: | 16 bit |
Dark noise: | 50 RMS counts |
Dynamic range: | 8.5 x 10^7 (system); 1300:1 for a single acquisition |
Integration time: | 1 ms – 65 seconds |
Stray light: | <0.05% at 600 nm; <0.10% at 435 nm |
Corrected linearity: | >99% |
ELECTRONICS | |
Power consumption: | 250 mA @ 5 VDC |
Inputs/Outputs: | Yes. Onboard digital user programmable GPIOS |
Trigger modes: | 4 modes |
Stobe functions: | Yes |
Gated delay feature: | Yes |
Connector: | 22-pin connector |
首先要注意到的是“Detector Range”,200~1100nm,
也就是可以量測的光譜範圍,
如果是深紫外光,例如EUV的172nm,那這個光譜儀就量不到,
遠紅外線到1200nm的也量不到,
不過對一般應用而言,這個範圍也很夠用了,
但是這裡的光譜量測範圍指得是感測器(Sony ILX 511B)能夠轉換為電訊號的光波長範圍,
並不是說這個光譜儀可以量測200-1100nm的光強度。
光譜量測範圍還跟分光系統有關,
舉例來說使用菱鏡分出來的色光只有400-700會投射在如下圖的28.672mm長度的這一段感測器有效像素上 :
那光譜儀可以量測的範圍就是400-700,其他波長範圍的光沒有投在檢測器上當然量不到。
第二個要注意的是“Optical resolution”,
0.1~10 nm FWHM Configuration dependent,
意思是能夠分辨的光波長值最精細可以到0.1~10nm之間。
特別的是補了一段文字 : “configuration dependent”,
也就是解析度是可以option設定出來的,
但是也代表可以作到的解析度需要跟其他的條件搭配才能作到。
要注意的是該規格後有註明了"FWHM",
FWHM在這裏的意思指得是光分離後進到每個Detector Sensor的Pixels上是一段分佈的連續波長,
例如透過光柵分光後,將波長700~1000投射在長度2048個Pixel的感測器上,
每個像素感測器等於分配到(1000-700)/2048 = 0.146nm的連續波長的光進入;
例如輸出結果是700.073nm的強度20000,
表示的是從700nm~700.146nm進入感測器像素後反映出的值,
所以想看到700.05nm跟700.10nm的光強度差異是不可能的事。
解析度跟光波長可量測範圍都會跟光譜儀的光路設計有關,包括狹縫的尺寸 、選用的光柵 、反射光的角度等等。
因此在選用光譜儀上首重要釐清楚光波長量測範圍,再來決定解析度要多高,其他的重點包括穩定性 、動態範圍表現 、暗訊號的抑制等等。